红外线心率计

发布时间:2013-06-11 12:40:32   来源:文档文库   
字号:

目录

1 .产品简介···········································1

2.主要芯片介绍········································1

2.1 CD4553··········································1

2.2 NE555···········································3

2.3 LM741···········································5

3 .红外线心率计工作原理·······························10

3.1 红外线心率计的原理框图···························10

3.2 单元电路的工作原理·······························10

1 血液波动检测电路···············································10

2 放大、整形、滤波电路·········································· 11

3 门控电路····················································· 12

4 3位计数电路··················································14

5 译码、驱动、显示电路·········································· 15

4 红外线心率计调试工艺·······························18

4.1 基本要求········································18

4.2 调试技术的发展···································18

4.3 红外线心率计的调试过程····························18

5.小结·················································21

6.参考文献·············································22

红外线心率计

1 .产品简介

红外线心率计就是通过红外线传感器检测出手指中动脉血管的微弱波动,由计数器计算出每分钟波动的次数。但手指中的毛细血管的波动是很微弱的,因此需要一个高放大倍数且低噪声的放大器,这是红外线心率计的设计关键所在。通过本产品的制作,可以使学生掌握常用模拟、数字集成电路(运算放大器、非门、555定时器、计数器、译码器等)的应用。

2.主要芯片介绍

2.1 cd4553

CD4553/CC45533位十进制计数器,但只有1个输出端,要完成3位输出,采用扫描输出方式,通过它的选通脉冲信号,依次控制3位十进制的输出,从而实现扫描显示方式。hUu838电子-技术资料-电子元件-电路图-技术应用网站-基本知识-原理-维修-作用-参数-电子元器件符号

引脚功能:hUu838电子-技术资料-电子元件-电路图-技术应用网站-基本知识-原理-维修-作用-参数-电子元器件符号

CLOCK:计数脉冲输入端,下调沿有效。   hUu838电子-技术资料-电子元件-电路图-技术应用网站-基本知识-原理-维修-作用-参数-电子元器件符号
CIACIB:内部振荡器的外界电容端子。   hUu838电子-技术资料-电子元件-电路图-技术应用网站-基本知识-原理-维修-作用-参数-电子元器件符号
MR:计数器清零(只清计数器部分),高电平有效。   hUu838电子-技术资料-电子元件-电路图-技术应用网站-基本知识-原理-维修-作用-参数-电子元器件符号
LE:锁定允许。当该端为低电平时,3组计数器的内容分别进入3组锁存器,当该端为高电平时,锁存器锁定,计数器的值不能进入。    hUu838电子-技术资料-电子元件-电路图-技术应用网站-基本知识-原理-维修-作用-参数-电子元器件符号
DIS:该端接地时,计数脉冲才能进行计数。   hUu838电子-技术资料-电子元件-电路图-技术应用网站-基本知识-原理-维修-作用-参数-电子元器件符号
DS1DS2DS3:位选通扫描信号的输出,这3端能循环地输出低电平,供显示器作为位通控制。   hUu838电子-技术资料-电子元件-电路图-技术应用网站-基本知识-原理-维修-作用-参数-电子元器件符号
Q0Q1Q2Q3BCD码输出端,它能分时轮流输出3组锁存器的BCD码。   hUu838电子-技术资料-电子元件-电路图-技术应用网站-基本知识-原理-维修-作用-参数-电子元器件符号
CD4553内部虽然有3BCD码计数器(计数最大值为999),但BCD的输出端却只有一组Q0Q3通过内部的多路转换开关能分时输出个、十、百位的BCD码,相应地,也输出3位位选通信号。例如:当Q0Q3输出个位的BCD码时,DS1端输出低电平;当Q0Q3输出十位的BCD码时,DS2端输出低电平;当Q0Q3输出百位的BCD码时,DS3端输出低电平时,周而复始、循环不止。hUu838电子-技术资料-电子元件-电路图-技术应用网站-基本知识-原理-维修-作用-参数-电子元器件符号

hUu838电子-技术资料-电子元件-电路图-技术应用网站-基本知识-原理-维修-作用-参数-电子元器件符号hUu838电子-技术资料-电子元件-电路图-技术应用网站-基本知识-原理-维修-作用-参数-电子元器件符号

              1  CD4553引脚hUu838电子-技术资料-电子元件-电路图-技术应用网站-基本知识-原理-维修-作用-参数-电子元器件符号

2.2 NE555

555芯片引脚图及引脚描述

5558脚是集成电路工作电压输入端,电压为518V,以UCC表示;从分压器上看出,上比较器6A1的5脚接在R1R2之间,所以5脚的电压固定在2UCC/3上;下比较器A2接在R2R3之间,A2的同相输入端电位被固定在UCC/3上。

NE555管脚功能介绍:

  1脚为地。2脚为触发输入端;3脚为输出端,输出的电平状态受触发器控制,而触发器受上比较器6脚和下比较器2脚的控制。

  当触发器接受上比较器A1R脚输入的高电平时,触发器被置于复位状态,3脚输出低电平;

  2脚和6脚是互补的,2脚只对低电平起作用,高电平对它不起作用,即电压小于1Ucc/3,此时3脚输出高电平。6脚为阈值端,只对高电平起作用,低电平对它不起作用,即输入 电压大于2 Ucc/3,称高触发端,3脚输出低电平,但有一个先决条件,即2脚电位必须大于1Ucc/3时才有效。3脚在高电位接近电源电压Ucc,输出电流最大可打200mA

  4脚是复位端,当4脚电位小于0.4V时,不管26脚状态如何,输出端3脚都输出低电平。

  5脚是控制端。

7脚称放电端,与3脚输出同步,输出电平一致,但7脚并不输出电流,所以3脚称为实高(或低)、7脚称为虚高。

1 555集成电路的框图及工作原理

555集成电路开始是作定时器应用的,所以叫做555定时器或555时基电路。但后来经过开发,它除了作定时延时控制外,还可用于调光、调温、调压、调速等多种控制及计量检测。此外,还可以组成脉冲振荡、单稳、双稳和脉冲调制电路,用于交流信号源、电源变换、频率变换、脉冲调制等。由于它工作可靠、使用方便、价格低廉,目前被广泛用于各种电子产品中,555集成电路内部有几十个元器件,有分压器、比较器、基本R-S触发器、放电管以及缓冲器等,电路比较复杂,是模拟电路和数字电路的混合体,如图1所示。

555芯片管脚介绍

555集成电路是8脚封装,双列直插型,如图2(A)所示,按输入输出的排列可看成如图2(B)所示。其中6脚称阈值端(TH),是上比较器的输入;2脚称触发端(TR),是下比较器的输入;3脚是输出端(Vo),它有O1两种状态,由输入端所加的电平决定;7脚是放电端(DIS),它是内部放电管的输出,有悬空和接地两种状态,也是由输入端的状态决定;4脚是复位端(MR),加上低电平时可使输出为低电平;5脚是控制电压端(Vc),可用它改变上下触发电平值;8脚是电源端,1脚是地端。

2 555集成电路封装图

我们也可以把555电路等效成一个带放电开关的R-S触发器,如图3(A)所示,这个特殊的触发器有两个输入端:阈值端(TH)可看成是置零端R,要求高电平,触发端(TR)可看成是置位端S,要求低电平,有一个输出端VoVo可等效成触发器的Q端,放电端(DIS)可看成是由内部放电开关控制的一个接点,由触发器的Q端控制:Q=1DIS端接地,Q=0DIS端悬空。另外还有复位端MR,控制电压端Vc,电源端VDD

地端GND。这个特殊的触发器有两个特点:

(1)两个输入端的触发电平要求一高一低,置零端R即阈值端(TH)要求高电平,而置位端s即触发端(TR)则要求低电乎;

(2)两个输入端的触发电平使输出发生翻转的阈值电压值也不同,当V c端不接控制电压时,对TH(R)端来讲,>2/3VDD是高电平1,<2/3VDD是低电平0:而对TR(S)端来讲,>1/3VDD是高电平1<1/3VDD是低电平0。如果在控制端(Vc)上控制电压Vc时,这时上触发电平就变成Vc值,下触发电平就变成1/2Vc值,可 见改变控制端的控制电压值就可以改变上下触发电平值。它的功能表见图3(B)所示。

3 555电路等效R—S触发器

555集成电路有双极型和CMOS型两种。CMOS型的优点是功耗低、电源电压低、输入阻抗高,但输出功率较小,输出驱动电流只有几毫安。双极型的优点是输出功率大,驱动电流达200毫安,其他指标则不如CMOS型的。

4. 典型应用电路

555的应用电路很多,只要改变555集成电路的外部附加电路,就可以构成几百种应用电路,大体上可分为555单稳、555双稳及555无稳(即振荡器)三类。

2.3 LM741

LM741/UA741运算放大器使用说明及应用

物理量的感测在一般应用中,经常使用各类传感器将位移、角度、压力、与流量等物理量转换为电流或电压信号,之后再由量测此电压电流信号间接推算出物理量变化,以达成感测、控制的目的。但有时传感器所输出的电压电流信号可能非常微小,以致信号处理时难以察觉其间的变化,故需要以放大器进行信号放大以顺利测得电流电压信号,而放大器所能达成的工作不仅是放大信号而已,尚能应用于缓冲隔离、准位转换、阻抗匹配、以及将电压转换为电流或电流转换为电压等用途。现今放大器种类繁多,一般仍以运算放大器(Operational Amplifier, Op Amp)应用较为广泛,本文即针对741运算放大器的使用加以说明。
1. 运算放大器简介ab126计算公式大全
放大器最初被开发的目的是运用于类比计算器之运算电路,其内部为复杂的集成电路(Integrated Circuit, IC),亦即在单一电子组件中整合了许多晶体管与二极管,图1为一般放大器之内部等值电路。


1. 运算放大器内部等值电路图
运算放大器属于使用反馈电路进行运算的高放大倍率型放大器,其放大倍率完全由外界组件所控制,透过外接电路或电阻的搭配,即可决定增益(即放大倍率)大小。图2为运算放大器于电路中的表示符号,可看出其包含两个输入端,其中(+)端为非反相(Non-Inverting)端,而(-)端称为反相(Inverting)端,运算放大器的作动与此二输入端差值有关,此差值称为「差动输入」。通常放大器的理想增益为无穷大,实际使用时亦往往相当高(可放大至105106倍),故差动输入跟增益后输出比较起来几乎等于零。838电子



2. 差动运算放大器表示符号
2. 741运算放大器使用说明
2.1 作动方式与原理新艺图库
741放大器为运算放大器中最常被使用的一种,拥有反相向与非反相两输入端,由输入端输入欲被放大的电流或电压信号,经放大后由输出端输出。放大器作动时的最大特点为需要一对同样大小的正负电源,其值由±12Vdc±18Vdc不等,而一般使用±15Vdc的电压。741运算放大器的外型与接脚配置分别如图34所示。
741运算放大器使用时需于74脚位供应一对同等大小的正负电源电压+Vdc与-Vdc,一旦于23脚位即两输入端间有电压差存在,压差即会被放大于输出端,唯Op放大器具有一特色,其输出电压值决不会大于正电源电压+Vdc或小于负电源电压-Vdc,输入电压差经放大后若大于外接电源电压+Vdc至-Vdc之范围,其值会等于+Vdc或-Vdc,故一般运算放大器输出电压均具有如图5之特性曲线,输出电压于到达+Vdc和-Vdc后会呈现饱和现象。



3. 放大器输出入电压关系图

741运算放大器之基本动作如图4所示,若在非反相输入端输入电压,会于输出端得到被放大的同极性输出;若以相同电压信号在反相输入端输入,则会在输出端获得放大相同倍率后但呈逆极性之信号输出。而当对放大器两输入端同时输入电压时,则是以非反相输入端电压值(V1)减去反相输入端电压值(V2),可于输出端得到(V1V2)经过倍率放大后之输出。

4. 放大器基本输出入关系图

3. 运算放大器常用电路
实际使用运算放大器时,因各类传感器输出电压变化极大,放大后电压很难正好落在放大器输出电压范围内,且运算放大器输入电源电压有其限制之承受范围限制,故需在电路上变化或补正。另外放大器放大倍率亦不一定正好为所需倍率,故需外接不同阻值电阻来解决。

3 .红外线心率计工作原理

3.1 红外线心率计的原理框图

整机电路由-10V电源变换电路、血液波动检测电路、放大整形滤波电路、3位计数器电路、门控电路、译码驱动显示电路组成,如图1所示。

1 红外线心率计的原理框图

3.2 单元电路的工作原理

1 血液波动检测电路

血液波动检测电路首先通过红外光电传感器把血液中波动的成分检测出来,然后通过电容器耦合到放大器的输入端。如图4所示。

4 血液波动检测电路

TCRT5000红外光电传感器的检测方法:

首先用数字万用表的二极管档位正向压降测试控制端发射管(浅蓝色)的正、负极,将红黑表笔分别接发射管的两个引脚,正反各测一次,表头一次显示“0.970.9-1.1)”,一次显示溢出值“-1”,则显示0.97V的那次正确,红表笔接的是正极,黑表笔接的是负极。若两次都显示“1”,说明发射管内部开路,若两次都显示“0”发射管内不短路。然后再判断接收管的CE极和光电转换效率,方法如下:将发射管的正负极分别插入数字万用表hFENPN型的CE插孔,再将模拟万用表打到R×1kΩ档。红黑表笔分别接接收管的两个引脚,若表针不动,则红黑表笔对调,若表针向右偏转到15kΩ左右,则黑表笔所接管脚为C,红表笔所接管脚为E。此时,再用手指或白纸贴近两管上方,表针继续向右偏转至1kΩ以内,说明该红外光电断续器的光电转换效率高。

血液波动检测电路工作原理:TCRT5000是集红外线发射管、接收管为一体的器件,工作时把探头贴在手指上,力度要适中。红外线发射管发出的红外线穿过动脉血管经手指指骨反射回来,反射回来的信号强度随着血液流动的变化而变化,接收管把反射回来的光信号变成微弱的电信号,并通过C1耦合到放大器。

2 放大、整形、滤波电路

放大、整形、滤波电路是把传感器检测到的微弱电信号进行放大、整形、滤波,最后输出反映心跳频率的方波,如图5所示。其中LM741为高精度单运放电路,它们的引脚功能如图3 (b)所示。IC2IC3IC4都为LM741

5 信号放大、整形电路

因为传感器送来的信号幅度只有25毫伏,要放大到10V左右才能作为计数器的输入脉冲。因此放大倍数设计在4000倍左右。两级放大器都接成反相比例放大器的电路,经过两级放大、反相后的波形是跟输入波形同相、且放大了的波形。放大后的波形是一个交流信号。其中A1A2的供电方式是正负电源供电,电源为+12V-10V

A1A2与周围元件组成二级放大电路,放大倍数Auf为:

由于放大后的波形是一个交流信号,而计数器需要的是单方向的直流脉冲信号。所以经过V3检波后变成单方向的直流脉冲信号,并把检波后的信号送到RC两阶滤波电路,滤波电路的作用是滤除放大后的干扰信号。R9V4组成传感器工作指示电路,当传感器接收到心跳信号时,V4就会按心跳的强度而改变亮度,因此V4正常工作时是按心跳的频率闪烁。直流脉冲信号滤波后送入A3的同相输入端,反相输入端接一个固定的电平,A3是作为一个电压比较器来工作的,是单电源供电。当A33脚电压高于2脚电压的时候,6脚输出高电平;当A33脚电压低于2脚电压的时候,6脚输出低电平,所以A3输出一个反应心跳频率的方波信号

3 门控电路

555定时器是一种将模拟电路和数字电路集成于一体的电子器件,用它可以构成单稳态触发器、多谐振荡器和施密特触发器等多种电路。555定时器在工业控制、定时、检测、报警等方面有广泛应用。

555定时器内部电路及其电路功能如图6(a)(b)所示。555内部电路由基本RS触发器FF、比较器COMP1COMP2和场效应管V1组成(参见图6(a)。当555内部的COMP1反相输入端(-)的输入信号VR小于其同相输入端(+)的比较电压VCO)时,COMP1输出高电位,置触发器FF为低电平,即Q=0;当COMP2同相输入端(+)的输入信号大于其反相输入端(-的比较电压VCO/2(1/3VDD)时,COMP2输出高电位,置触发器FF为高电平,即Q=1是直接复位端,Q=0MOSV1是单稳态等定时电路时,供定时电容C对地放电作用。

注意:电压VCO可以外部提供,故称外加控制电压,也可以使用内部分压器产生的电压,这时COMP2的比较电压为VDD/3,不用时常接0.01μF电容到地以防干扰。

(a) 555定时器内部电路 (b) 555简化符号

6 555定时器内部电路及其功能符号

555接成单稳态触发器来完成门控电路的作用是控制计数器的启停,并控制每次测量的时间,电路如图7(a)所示。

(a) 电路 (b) 工作波形

7 555组成的门控电路

当接通电源的时候,+12V电源电压通过R15对电容C4进行充电,2脚的电压马上变成12V(“1电平),触发器FF被置“0 5553脚输出“0电平(参见图7(a))。V6截止,V6C极为高电位,所以计数器MC14553不计数,此时V5不亮。

当按下S1按钮时,2脚电压为0V,低于1/3电源电压。555内部CMP2输出高电平(参见图6(a),触发器FF被置“1,即3脚输出“1电平,V6饱和导通,V5发光,V6集电极输出低电平,使计数器MC14553清零,开始计数。同时555内场效应管截止,12V电压通过R17C6充电,C6的电压逐渐增高,如图7(b)uC6波形。

C6的电压充到2/3电源电压的时候,555CMP1输出高电平,触发器置“03脚输出低电平,V6集电极输出高电平,因此计数器MC1455311脚变为高电平,计数器停止计数;同时555内场效应管导通,电容C6通过场效应管迅速放电到低电平,返回稳定的状态,定时结束。

脉宽TW可根据下式计算:

1-1

4 3位计数电路

MC14553组成的3位计数电路对输入的方波进行计数,并把计数结果以BCD码的形式输出。

MC14553为十六引脚扁平封装集成电路,其引脚功能如图8a)所示,有四个BCD码输出端Q1Q3,可分时输出三组BCD码;有三个分时同步控制信号DS1DS3,为计数器的输出提供分时同步输出控制信号,形成动态扫描工作方式,该控制端低电平有效。计数电路包含了计数和输出驱动电路。

(a) MC14553 (b) CD4543

8 集成电路引脚功能图

计数器MC14553真值表如下:

2.1 MC14553真值表

X=任意

计数器MC14553 DS1DS3输出为方波,波形如图8所示。当按下S1时(参见图7(a)V5饱和导通,V5C极为低电平,MC1455311脚变为低电平,计数器开始对送到12脚的从整形电路过来的方波个数进行计数,最大计数为999,计数结果以BCD码的形式从Q0Q3输出。11脚不管是高电平还是低电平,DS1DS3始终是输出图9的方波。当DS3是低电平的时候,个位显示器被选中,Q0Q3输出个位要显示的数值;当DS2是低电平的时候,十位显示器被选中,Q0Q3输出十位要显示的数值;当DS1是低电平的时候,百位显示器被选中,Q0Q3输出百位要显示的数值。

9 DS1DS3输出波形图

5 译码、驱动、显示电路

3位计数电路、译码、驱动、显示电路如图10所示,它的作用是把计数器输出的计数结果显示在3位数码管上。

译码器CD4543的引脚功能如图8b)所示。它有了四个输入端:ABCD,与计数器的输出端相连;有七个数码笔段输出驱动端:ag。译码器CD4543可以驱动共阴、共阳两种数码管,使用时,只要将PH引脚接高电平,即可驱动共阳极的LED数码管;将PH引脚接低电平,即可驱动共阴极的LED数码管。

显示采取动态扫描的方法,即每一时刻只有一个数码管被点亮,但是交替的频率非常快,由于人眼的视觉残留效应,人眼看到的就是静止的数字显示结果。计数器送来的数据,经过CD4543翻译成7段字码后,接到数码管的7个笔画端,点亮相应的笔画段。数码管采用共阳极的。CD4543的真值表如下:

2.2 CD4543的真值表

10 3位计数、译码、驱动、显示电路

上图中:IC6MC14553IC7CD4543

4 .红外线心率计调试工艺

4.1 基本要求

熟练掌握用数字万用表测量集成电路以及各种元器件引脚的电压,判别各种集成电路输入、输出状态;熟练使用示波器测量要求点的电压波形。

会利用电原理分析和排除调试过程中出现的故障。

4.2 调试技术的发展:

当代电子产品调试技术的发展有三个明显的趋势:

强调整体调试

由于微电子技术和EDA(电子设计自动化)技术飞速发展,电子产品元器件数量减少,设计制造水平不断提高,使得大批产品内的同一功能电路之间差别微小,不同工能模块之间的配合有条件在产品设计中解决,因此一个电子产品内各功能块不需要或很少需要调试。调试工作主要集中在整体调试,极大提高了效率,降低了产品制造成本。

趋向免调试、少测试

由于现在很多电子产品采用高集成度专用的集成电路;大规模、超大规模通用集成电路和高质量电子元器件组成,制造工艺先进,使电子产品调试走出传统反复调整和测试的模式,向免调试、少测试方向发展。例如现代采用单片集成电路和SMT技术制造的数字调谐收音机,几乎不用调整,只须很少测试便可以达到较高的指标。

发展自动测试

先进的计算机集成制造系统将电子产品的测试完全由计算机控制,产品的一致性和质量都达到空前的水平。

4.3 红外线心率计的调试过程

电源变换电路的调试

12V直流电压送入电源变换电路CD40961脚(正极)与8脚(负极)之间,注意正负极。用数字万用表DC20V挡测集成电路CD40961脚与12脚之间电压为12V。此时,RC振荡器应该工作,用示波器(量程DC5V/DIV)测试C点(参见图3)的电压波形,应该是一个不规范的方波,测试D点的波形,应该是规范的方波,把C点、D点测量电压波形记录在表4.1中。用数字万用表的DC20V挡测量E点对地的电压,应该是-10V左右,把测得电压记录在记录表3中。(如果没有方波,说明电路没有起振,应检查RC电路和芯片4096的电源有没有接错;如果电路起振,而没有-10V电压,检查二极管V1V2和电解电容C6的极性有没有接错。

血液波动检测电路的调试

电路连接完毕后通电。把食指放在传感器的探头处,适当调节压力。用示波器的AC挡(5mV/DIV500mS/DIV)去测量电容C1正极对地的波形,应该能在示波器上看到心跳微弱的波动,幅度大约是几个毫伏。如果有此波形,说明传感器工作正常,如果没有,检查传感器的引脚是否接错。用示波器去测量C1负极的波形,应该能看到比刚才那个波形幅度还要小的波动。

放大、整形、滤波电路的调试

电路连接完毕后通电。测量IC2IC37脚、4脚对直流地的电压(即运放的供电电压),应该为+12V -10V左右。测量IC47脚、4脚之间的电压,应该为+12V左右。

把食指放在传感器(ON2152)的探头上,适当调节压力。V4应该会有节律的闪烁,闪烁的频率跟心跳的频率吻合。此时,用示波器测量IC26脚波形,应该是放大了R4/R3倍的波动信号。用示波器测量IC36脚波形,应该是比IC26脚波形放大了R8/R6倍的波形(因为放大倍数很大,波形有削顶现象)IC2IC3的放大倍数可以根据自己的实际情况适当做一些调整。用示波器测量IC46脚波形,应该是一个规范的方波,是单极性的,如果没有方波,或方波的占空比太小,可以适当改变R10R11的阻值。把测得的三个波形画入表4.3中。

如果手上暂时没有传感器,则使用函数信号发生器产生几hz、几mv的正弦波,并把该波形加到C1的负极,同样也可以按上述的方法进行调试。

门控电路的调试

电路连接完毕后通电。此时,门控电路进入稳态,用数字万用表DC20V挡测量367脚与1脚之间的电压都为0VV6C极与1脚之间的电压为12VV5不发光。按一下S1按钮,门控电路输出状态发生翻转,进入暂稳态,555输出端3脚输出高电位,因此V6饱和导通,V6C极输出低电位,V5发光,用数字万用表DC20V挡测量67脚与1脚之间的电压,可以发现,电压是慢慢上升的,当上升到8V左右的时候(时间是30秒),门控电路输出状态又发生翻转,进入稳态,此时555输出端3脚输出低电位。用数字万用表DC20V挡测量367脚的与1脚之间的电压,都是0VV6C极与1脚之间的电压为12VV5不发光。

如果暂稳态的时间不是30秒,则最后测量的心率不准确。需要调整R17C6的参数来达到30秒的要求。具体计算公式:1.1× R17× C6=30

计数、译码、驱动、显示电路的调试

电路连接完毕后通电。此时由于门控电路的控制作用,计数器MC14553的使能端(低电平有效)被置“1”,计数器不计数,输出的BCD码是00005679脚的电压大约是0V。用示波器双踪测量DS1DS2之间、DS2DS3之间波形,应能显示图(8)所示的波形,测试并把波形画在4.5中(示波器量程:双踪,5V/DIV1mS/DIV)。

把食指放在传感器的探头处,适当调节压力。当观察到V4呈现有规律的亮-灭时,就可以进行测量了。按一下门控电路的S1,这时,V5发光,计数器的使能端被置“0”,计数器开始按整形电路送来的心跳脉冲计数。计数的结果以BCD码的形式送到译码器进行译码。译码后的结果送到数码管显示计数的结果。过30秒钟后,门控电路输出高电平,计数器使能端被置“1”,计数器停止计数。数码管显示最后计数的结果,此数字乘2即是被测的心率。测量并记录计数器停止计数后,集成电路MC14553CD4543的引脚电压并填入表4.6、表4.7

附件1:元器件清单

5.小结

红外线心率计就是通过红外线传感器检测出手指中动脉血管的微弱波动,由计数器计算出每分钟波动的次数。但手指中的毛细血管的波动是很微弱的,因此需要一个高放大倍数且低噪声的放大器,这是红外线心率计的设计关键所在。

心率是指单位时间内心脏波动的次数。一般指每分钟的心跳次数,是临床常规检查的指标。传统测量脉搏次数的方法是以基准时间为单位,并通过基准时间内测得的脉搏跳动次数和相关倍数,计算每分钟的脉搏次数。但在现实生活中,这种可以直接测量心跳的一起还不是很常用,在很多小型医院里,医生们依旧使用着古老的手动式听诊器,这大大影响了效率和救治时间。这就为心率计的研究与生产奠定了必然基础。

待方案确定以后,检测与修改是必不可少的。修正设计方案的实践过程,也是应用理论知识来解决实践中各类问题的关键环节,是电路设计者必须掌握的基本技能。

把电子元器件连接起来,实现特定功能的关键一步是调试。调试方法有两种:分块调试发和整体调试法。

在这次红外线心率计的课程设计中我大体掌握常用模拟、数字集成电路(运算放大器、非门、555定时器、计数器、译码器等)的应用。同时也遇到了相当多的困难,在和同学的讨论以及老师的知道下,都一一解决了。

6.参考文献

本文来源:https://www.2haoxitong.net/k/doc/d514de3ee2bd960591c67709.html

《红外线心率计.doc》
将本文的Word文档下载到电脑,方便收藏和打印
推荐度:
点击下载文档

文档为doc格式